Demodulator calibration procedure of the ProxiLAB in reader mode

PCD assembly with active bridge (External)

*Configure the test bench with the Test PCD Assembly and the amplifier connected

*Put a sample in DUT position.

*Open RGPA, tab "Settings\General"

*In Demodulator calibratio, check "External" (Reader)

*Click on "Run calibration"

NB: If you want to use the SSB, please turn on both switches on the bridge and check the SSB box in RGPA before the calibration. If you deactivate the SSB, please run the calibration again.

 

Rf In demodulator

*Connect a Poller 0 to the ANT plug

*Plug the probe you'll use to the PROBE plug

*Put the probe on the antenna

*Open RGPA, tab "Settings\General"

*In Demodulator calibratio, check "Probe" (Reader)

*Click on "Run calibration"

NB: If you're using a Test PCD Assembly, you can also set up the bench with the amplifier and PICC on DUT position. Then, put the probe on the sample and run calibration.