4.46.21.0909

Product: 

Improvements

  • Self-control mechanism during the Power-up sequence added
  • ISO14443: reader and emulator handle FSC/FSD up to 4096 bytes (256 bytes before)
  • JISX6319 : PICC demodulation
  • ISO15693 : PICC and PCD decoding
  • ISO18092 : ProxiLAB should be able to send a RSL_REQ without DID field
  • Spy logger handle frames up to 4096 bytes (1024 bytes before)
  • Power level can now be specified in percent for more precision
  • A residual power field can be specified when RF is off
  • Maximum waveform overshoot is up to 200% of high level (125% of low/high level before)
  • API functions for ALM (Active Load Modulation) and subcarriers
  • VNA : reduced calibration save and reload time, new enable/disable port extension API

 Fixes

  • ISO15693: TX logger in emulator mode (first byte of PICC frame not logged)
  • AWG noisy output (due to DAC configuration)
  • JISX6319: Some Write commands were analyzed as UNDECODED FRAME even though they were correctly formatted
  • Administrator mode is not required anymore to update ProxiLAB licenses with a .key file
  • Some ISO14443 Type B frames were mistakenly decoded as Innovatron (B prime) frames
  • ISO14443-B analyzer : analysis of ATTRIB command will now display correct warning messages
  • ISO14443-B analyzer : glitch under 20us are ignored
  • ISO18092 analyzer : the warning "Bits b4-b7 shall be set to 0, all other values are RFU" was wrongly displayed for ATR_RES frames
  • ISO18092 : bitrate management fix
  • APDU : details now available in description view
  • S-Param : correctly displayed in description view
  • ISO15693 : PCD frame length should not vary with pause length
  • ISO15693 : fixed CRC computation in description view
  • ISO14443 : the warning "Bit b4 of PCB is set but CID is not present" was wrongly displayed

 

Known limitations:

  • This version of ProxiLAB software does not support PCSC features

 

Before updating your equipment, and due to the self-control mechanism during the power-up sequence, make sure you have returned your equipment for the annual control & calibration. Please contact our support team for more information support@keolabs.com.