Demodulator calibration procedure of the ProxiLAB in PICC emulator mode

1.       To calibrate the ProxiLAB in PICC emulator mode using the SC-ProxiLAB-PICC-Probe+ follow this procedure:

Hardware configuration:

o    Connect SC-ProxiLAB-PICC-Probe+ on the ProxiLAB (EXT.) using a HDMI cable.

o    Connect the demodulator probe (CALIBRATION COIL 2, PCD Probe 1 or other probes) to the SC-ProxiLAB-PICC-Probe+ (J7 RF) using a SMA-SMA cable.

o    Connect the SC-ProxiLAB-Poller0 (J1 IN) on the ProxiLAB output (ANT.) using a SMA-SMA cable.

o    Place the demodulator probe on the SC-ProxiLAB-Poller0.

 

In RGPA software:

o    Connect the ProxiLAB tool.

o    In Settings->General : select PICC Emulator (Type A, B) in Smart card interface source section, then click on Default in Pre-Configuration section.

o    In Settings->Emulator : uncheck Type B.

o    In Settings->General : Select External - Probe in Demodulator calibration section as shown on the Picture 1.

o    Click on Run calibration and click OK on the pop-ups.

 

Hardware configuration:

o    Disconnect the demodulator probe from the SC-ProxiLAB-PICC-Probe+ (J7 RF) 

o    Connect the demodulator probe on the ProxiLAB input (PROBE) using a SMA-SMA cable.

The ProxiLAB is ready to be used in the test suite.

2.       To calibrate the ProxiLAB in PICC emulator mode using the SC-ProxiLAB PICC-Probe follow this procedure:

Hardware configuration:

o    Connect SC-ProxiLAB PICC-Probe on the ProxiLAB (EXT.) using a HDMI cable.

o    Connect the SC-ProxiLAB-Poller0 (J1 IN) on the ProxiLAB output (ANT.) using a SMA-SMA cable.

o    Place the SC-ProxiLAB PICC-Probe on the SC-ProxiLAB-Poller0.

 

In RGPA software:

o    Connect the ProxiLAB tool.

o    In Settings->General : select PICC Emulator (Type A, B) in Smart card interface source section, then click on Default in Pre-Configuration section.

o    In Settings->Emulator : uncheck Type B.

o    In Settings->General : Select External - Probe in Demodulator calibration section as shown on the Picture 1.

o    Click on Run calibration and click OK on the pop-ups.

 

The ProxiLAB is ready to be used in the test suite.

 

Picture 1:

Demodulator calibration